Part
III: Culture
quality control
Quality
assurance with regard to tuberculosis culture is a system
designed to continuously improve the reliability, efficiency
and use of culture as diagnostic and monitoring option. The
purpose of a quality assurance programme is to improve the efficiency
and reliability of culture services. The components of a quality
assurance programme are:
- quality
control
- quality
improvement
- proficiency
testing
The
following section will focus on aspects of quality controlin
the culture laboratory. For a discussion on quality improvement
and proficiency testing please refer to the Management Series.
Quality
control of culture is a process of effective and systematic
internal monitoring of the performance of bench work
in the culture laboratory. Quality control ensures that the
information generated by the laboratory is accurate, reliable
and reproducible. This is accomplished by assessing - against
acceptable established limits - the quality of specimens, the
performance of decontamination, digestion and culture procedures,
the quality of reagents, media and equipment, by reviewing culture
results and by documenting the validity of culture methods.
Quality
control should be performed on a regular basis in the culture
laboratory to ensure reliability and reproducibility of laboratory
results. For a quality control programme to be of value, it
must be practical and workable.
|
Quality
control is the responsibility of all laboratory workers |
Quality
control must be applied to:
- laboratory
arrangement
- equipment
- collection
and transport of specimens
- handling
of specimens
- reagents
and media
- culture
methods
- reporting
of results
The
keys to successful quality control are:
- adequately
trained, interested and committed staff
- common-sense
use of practical procedures
- a
willingness to admit and rectify mistakes
- effective
communication
Quality
control measures which must be in place in all tuberculosis
culture laboratories include:
Laboratory arrangement
and administration
- Ensure
that doors in the laboratory are always closed. Work
areas, equipment and supplies should be arranged for logical
and efficient work flow. Work areas should be kept free of
dust. Benches should be swabbed at least once a day with an
appropriate disinfectant (eg. 5% phenol)
- Every procedure
performed in the laboratory must be written out exactly as
carried out and be kept in the laboratory for easy reference.
Any changes must be dated and initialised by the laboratory
supervisor
- All
records should be retained for two years
- Laboratory
procedures used routinely should be those that have been published
in reputable microbiological books, manuals or journals
Laboratory
equipment
- Equipment
should meet the manufacturers claims and specifications
- Written
operating and cleaning instructions must be kept in a file
for all equipment
- Dated
service records must be kept for all equipment
- Equipment
must be monitored regularly to ensure the constant accuracy
and precision necessary.
- Biological
safety cabinet: The BSC is the
primary containment device that protects the worker, product
and/or environment from exposure to tuberculosis and its
performance needs to be verified at the time of installation
and annually thereafter. The purpose and acceptance
level of the performance tests are to ensure the balance
of inflow and exhaust air, the distribution of air
onto the work surface and the integrity of the cabinet.
Other tests check electrical and physical features
of the BSC.
Daily
checks of the BSC include the following:
- Ensure
that the rate of airflow across the front opening is 75 linear
feet/minute (22.86 meter/second) for Class I and 75
to 100 linear feet/minute (22.86 to 30.48 meter/second) for
Class II cabinets.
- Check
the magnetic gauge in the exhaust duct for any pressure drop
across the filters and replace the filters when the
gauge indicates that the airflow across the front opening
has dropped below optimal levels.
The
following tests should be performed annually on Class
I and Class II cabinets:
Downflow
velocity and volume test
This
test is performed to measure the velocity of air moving
through the cabinet workspace.
This test is performed
to determine the calculated or directly measured velocity
through the work access opening, to verify the nominal set point
average inflow velocity and to calculate the exhaust airflow
volume rate.
An
electronic vane type anemometer should be used to measure airflow.
The airflow into a Class I cabinet should be measured in at
least five places in the plane of the working face and
an average calculated. At no place should there be a reading
that is 20 linear feet/minute (0.1meter/second) more or less
than any of the others. If there is such a difference
there will be turbulence within the cabinet.
In
Class II cabinets the airflow is greater at the bottom than
at the top of the working face. The average inward flow
is calculated by measuring the velocity of air leaving the exhaust
and the area of the exhaust vent. From this the volume per minute
is calculated, which is also the amount entering the cabinet.
Divided by the area of the working face it gives the average
velocity. The downward velocity of air should be measured at
18 points in the horizontal place, 10cm above the top edge of
the working face. No reading should differ from the mean by
more than 20%.
- Airflow
smoke patterns tests
This
test is performed to determine if the airflow along the entire
perimeter of the work access opening is inward, if airflow within
the work area is downward with no dead spots or refluxing, if
ambient air passes onto or over the work surface, and
if there is refluxing to the outside at the window wiper gasket
and side seals. The smoke test is an indicator of airflow direction,
not of velocity.
Commercial
airflow testers are recommended. They are small glass tubes,
sealed at each end. Both ends are broken off with the gadget
provided and a rubber bulb fitted to one end. Pressing the bulb
to pass air through the tube causes it to emit white smoke.
This
test is performed to determine the integrity of supply
and exhaust HEPA filters, filter housing, and after-mounting
frames while the cabinet is operated at the nominal set
point velocities. An aerosol in the form of generated
particulates of dioctylphthalate (DOP) or an accepted alternative
is required for leak-testing HEPA filters and their seals.
Although DOP has been identified as a potential carcinogen,
competent service personnel are trained to use this chemical
in a safe manner. The aerosol is generated on the intake
side of the filter, and particles passing through the filter
or around the seal are measured with a photometer on the
discharge side. This test is suitable for ascertaining the integrity
of all HEPA filters.
The
pressure holding test is performed to determine if exterior
surfaces of all plenums, welds, gaskets, and plenum penetrations
or seals are free of leaks. It is performed just prior to initial
installation when the BSC is in a free-standing position in
the room in which it will be used, after a cabinet has
been relocated to a new location, and again after removal of
access panels to plenums for repairs or a filter change. This
test may also be performed on fully installed cabinets.
- Electrical leakage and ground
circuit resistance and polarity tests
These
safety tests are performed to determine if a potential
shock hazard exists by measuring the electrical leakage, polarity
ground fault interrupter function, and ground circuit resistance
to the cabinet connection. The polarity of electrical outlets
are checked using a polarity tester. The ground fault
circuit interrupter should trip when approximately 5 milliampere
(ma) is applied.
This
test is performed to measure the light intensity on the
work surface of the cabinet as an aid in minimising cabinet
operator?s
fatigue.
This
test is performed to determine the amount of vibration
in an operating cabinet as a guide to satisfactory mechanical
performance, as an aid in minimising cabinet operator?s
fatigue, and to prevent damage to delicate tissue culture specimens.
This test is performed to
measure the noise levels produced by the cabinet, as a
guide to satisfactory mechanical performance and an aid
in minimising cabinet operator?s
fatigue.
- Centrifuge: Check brushes and bearings
every 6 months
- Incubator
35E-37EC:
Record the temperature daily,
preferably in the morning. Test the temperature at several
sites within the incubator by placing a thermometer
in a water reservoir (eg. Erlenmeyer flask). Control the light
within the incubator by covering the glass front of the incubator
door and by restricting the use of any lights inside
the incubator
- Inspissator: Check temperature
daily. Clean after each batch of culture media prepared
- pH
meter:
Compensate for temperature with each run. Date buffer solutions
and discard when unsatisfactory. Standardise with pH 4.0 and
7.0 buffers before each test or series of tests
- Water
baths:Check
temperature before and during use. Clean monthly
-
- Refrigerator
2E-8EC:
Check
temperature daily. Clean monthly. Defrost or check refrigerator
and freezer compartment every 3 months
- Freezers:
Check daily. Clean every 6 months
- Glassware:
Discard chipped or etched glassware. Ensure that glassware
are free of detergents. Do not store sterile glassware for
more than three weeks before it is used
Specimens
and request forms
- Perform
tests only upon written request of authorised persons
and do not allow oral requests without follow-up written instructions
- Insist
on specimen request forms being kept separate from the
specimens themselves. Forms that have been contaminated
by specimens should be sterilised by autoclaving
- Insist
on adequately completed request forms and proper labelling
of specimens to ensure positive identification of patients.
Reject specimens that cannot be properly identified
- Evaluate
the quality of sputum specimens and make a note if a specimen
resembles saliva. The report should state ?specimen
resembled saliva - treat a negative result with caution?
(to facilitate reporting a rubber stamp of the comment can
be made)
- Discard leaking and
broken specimen containers by autoclaving and request a repeat
specimen
- Document
the arrival time of specimens in the laboratory and note any
delays in delivery on the report form, particularly
with negative/contaminated results
Reagents
and stains
- All
containers of stains and reagents should show the date received
and the date first opened. Any material found to be
unsatisfactory should be recorded as such and removed
from the laboratory immediately. Stocks should be limited
to six months?
supply and regular stock rotation should take place to avoid
unnecessary expiry.
Digestion
and decontamination
- Process
sputum specimens in batches according to centrifuge capacity
- Keep
a monthly record of the percentage of clinical specimens contaminated:
the acceptable range is 2-5%. Contamination rates <2% indicates
overly harsh decontamination, which means that too many
tubercle bacilli are killed. If the laboratory is experiencing
delays in delivery of specimens the contamination rate may
be greater than 5%. If a rate of >5% persists, ensure
that specimens are completely digested, since partially digested
specimens may not be completely decontaminated. Thoroughly
mix the contents of centrifuge tubes to ensure that the inside
surfaces have been well decontaminated
Culture
media
- Use
fresh eggs (< seven days) for preparation of Löwenstein-Jensen
media
- Control
coagulation time and temperature for egg-based medium. Discard
media that are discoloured or have bubbles following inspissation
- Check
all batches of media for sterility by incubation at
35E-37EC
for 24 hours
- Keep all media
in the dark in the refrigerator and discard unused media after
four weeks
Culture
procedures
- Avoid
cross-contamination of cultures by using individual pipettes
or loops and strict aseptic techniques
- Be
suspicious of several successively positive specimens
or of cultures with few colonies that follow a heavily positive
culture
Biochemical
tests
Prepare
reagents as indicated and check the expected biochemical test
response by using appropriate positive and negative controls
Water
Check
both distilled and tap water regularly for the presence of acid-fast
contaminants. If water appears cloudy or dirty, centrifuge 200-250ml
in multiple tubes and make a smear of the combined sediment.
Alternatively, filter 1 000ml of water through a sterile 0.22Fm
pore size membrane filter, cut the filter into strips with a
sterile scissor and place on Löwenstein-Jensen culture medium.
|