Part
II: Microscopy
smear examination procedures
Because the human eye cannot
see objects with a diameter of less than 0.1mm, individual bacteria
can only be detected with a microscope. The function of the
magnifying lens system of the microscope is to magnify the objects
within the microscopic field to a size which can be detected
by the human eye. In addition to magnification, two other factors,
contrast and resolution, are of great importance. In order to
be perceived through the microscope, an object must possess
a certain degree of contrast with its surrounding medium: in
order to produce a clear magnified image, the microscope must
possess a resolving power sufficient to permit perception of
individual objects. The degree of contrast can be greatly increased
by staining procedures: treatment with dyes that bind selectively
either to the whole cell or to certain cell components. The
acid-fast stain is used to obtain information on the composition
of the cell walls of tubercle bacilli. The types of microscopy
most commonly used for observing acid-fast bacilli are bright-field
and fluorescence microscopy. In bright-field microscopy the
light passes through the bacilli and the variation in colour
due to staining show the form of the organisms. Fluorescence
stains are usually organic substances which absorb ultraviolet
light and remit part of the energy as light of longer wavelength
which can be observed through the eyepiece as fluorescence.
When exposed to ultraviolet light, the fluorescent bacilli are
perceived as brightly coloured organisms against a dark background.
A
binocular microscope (ie. one with two eye pieces) is recommended
for the examination of stained smears. If a monocular microscope
is used, both eyes should be kept open while looking into the
eyepiece, to prevent eye fatigue. If no electricity is available,
daylight must be used as light source and the table with the
microscope be placed in front of a window.
Components
of the microscope
The
components of the microscope are illustrated in Figure 3.
Mechanical
part
The
base or foot of the microscope should be sufficiently
heavy to be stable as it supports the stage on which
the slide is placed. The slide is kept steady by clips attached
to the stage; movements are guided by a micrometric screw with
vernier to help relocate a field of examination.
The
limb supports the eyepiece holder and the objective holder.
In the monocular microscope the two devices are separated by
a tube with the distance between the eyepiece and the objective
approximately 170mm. In binocular microscopes the tube is replaced
by a system of prisms.
In
the monocular model, the eyepiece holder is a simple tube, while
in the binocular model, the prisms are placed within the eyepiece
holder. The two eyepieces can be moved to adjust to the distance
between the observer?s
pupils; one of the eyepieces moves around its axis to correct
any difference in convergence between both eyes.
The objective holder
can be rotated as well as replaced, if necessary.
Optical
outfit
The
mirror, which has two faces, plane and concave, sends
parallel rays from the light source to the optical axis of the
microscope.
The
light source is usually incorporated in the foot of the
microscope. The fixed light source keeps a constant distance
and alignment. Usually small halogen bulbs of great intensity
are used. Tungsten filament lamps can also be employed.
The condenser is a
lens that focuses the light on the slide. It is centered by
moving the centering screws until the iris is concentric with
the opening of the back of the objective. The greater the aperture
of the diaphragm the wider the angle of vision. The light opening
must be smaller than the opening of the objective used.
Objectives
Objectives
are classified according to their numerical aperture (NA) and
magnification. The NA describes the properties of the objective
lens. Up to a certain limit an increase in the NA of the objective
lens increases resolving power.
The passage of light through
curved lenses produces the separation of the spectrum in the
different wave lengths. This results in coloured bands in the
periphery of the lens due to a chromatic aberration. The greater
the magnification, the greater the aberration will be. The different
colours (wave lengths) do not focus on the same place, producing
spherical aberration. These are compensated for by using a combination
of lenses. Achromatic objectives are corrected for two
colours of the chromatic aberration (yellow and green) and for
one of the spherical aberration. Apochromatic objectives
do not present chromatic aberrations throughout the whole scale
of visible spectrum provided there is an optimal distance between
the object and the lens of the objective. They are classified
by distance in mm or fractions of mm. They give clearer and
crisper images than the achromatic lenses and can be used with
higher power oculars. The flat apochromatic lenses have a correction
for spherical aberration.
Immersion
objectives
Dry
objectives are those where the front lens is in contact with
no other medium than the air (refraction index nil). The refraction
index of glass (as used for microscope slides) is 1.51. As a
result, light rays are refracted in passing from the glass to
the air; some fall beyond the visual scope of the objectives.
With oil-immersion objectives, the oil is used to fill the space
between the front lens and the slide. To achieve high magnification,
immersion oil is placed between the slide and the immersion
objective lens. Unlike air, immersion oil has the same refractive
index as glass and consequently improves the resolving power
of the lens.
The eyepiece
or ocular is composed of two lenses mounted on two ends of a
tube. The lower lens is called the collector or field glass;
it flattens and clears the real image and therefore completes
the action of the objectives. The upper lens enlarges the image
and therefore gives a virtual image of the object.
In
brief, the objective produces an image of the preparation inside
the microscope, and the observer looks at the image through
the eyepiece. To form a clear image the lens must focus each
point in the slide to give a point in the image.
Use
of the microscope
- Check for broken or damaged
parts
- Ensure
that the light source is well regulated and focused
- Ensure
that the lenses, mirrors and other light-conducting surfaces
are clean
- Ensure
that the condenser is in the upper position with the
diaphragm open
- Adjust
the light, mirror, condenser and diagram so that a strong
beam of light is directed towards the objective lens
- Turn
the coarse adjustment knob to move the objective lens away
from the stage
- Rotate the nosepiece so that
a low power objective lens (5x or 10x) is directly over
the condenser
- Place
the slide on the stage so that the smear is directly under
the objective lens
- Look
from the side of the stage to observe the space between the
slide and the objective lens. Slowly turn the coarse focus
knob to bring the objective lens close - but not touching
- the smear
- Adjust
the light intensity so that it is bright but not uncomfortable
when looking into the eyepiece. This may be done by changing
the intensity of the lamp, changing mirror surfaces,
using dark filters, adjusting the diaphragm or adjusting
the condenser. Usually, adjusting the diaphragm is sufficient
- While
looking into the eyepiece, slowly turn the coarse focus
knob to separate the objective lens and the stage. The smear
should come into focus within a few turns
- Turn
the fine focus knob until the smear is seen most clearly
- While looking from the side,
turn the nosepiece to select a higher power lens. Ensure
that the lens does not touch the slide. The smear should almost
be in focus. The best focus can be achieved by adjusting the
fine focus knob. A light adjustment may also be helpful
- To
use the oil immersion lens, turn the nosepiece so that
the lens is over the smear. Put a drop of immersion oil on
the smear - do not touch the slide with the oil applicator
but allow the drop of oil to fall freely onto the smear
- Lower
the 100x immersion lens so that it comes into contact
with the oil. Slowly bring the immersion lens upwards until
the image of the smear appears. Adjust the fine focus knob
to focus
Examination
procedures
To
obtain excellence in microscopic examination, a good microscope
and a comfortable work area is required. Reading of smears must
be systematic and standardised to ensure that a representative
area of the smear is examined. To ensure that an area is covered
only once, the smear should be examined in an orderly manner
and the following procedure is recommended:
- Always
examine carbolfuchsin-stained smears with a 100x oil immersion
objective
- Examine
fluorochrome-stained smears within 24 hours of staining as
the fluorescence may fade with time. Smears that cannot
be examined immediately after staining should be kept in the
dark, preferably in a refrigerator, for a maximum of 24 hours
- If
possible include a known positive slide and a known
negative slide with each day?s
work. The positive control ensures the staining capability
of the solutions and of the staining procedure. The
negative control confirms that acid-fast contaminants are
not present in the stains or in other solutions
- Make
a series of systematic sweeps over the length of the smear.
After examining a microscopic field, move the slide longitudinally
so that the neighbouring field to the right can be examined.
Search each field thoroughly
- Examine
a minimum of 100 fields before the smear is reported as negative.
For a skilled microscopist this will take approximately
five minutes. In a smear of 1.5cm x 1.5cm the number
of microscopic fields in one length of the slide corresponds
to around 100. If the smear is moderately or heavily positive
fewer fields may be examined and a report of ?positive?
may be made even though the entire smear has not been examined
- At the end of examination,
take the slide from the microscope stage, check the identification
number, and note the result. Dip the slide into xylene to
remove the immersion oil and place it in a box for examined
slides
- Before
examining the next slide, wipe the immersion lens with a piece
of lens tissue
- Unexpected objects may be
seen when using the microscope. If these objects move only
when the slide is moved, they may be materials occasionally
found in the specimen or object, precipitated fixatives
or stains, contaminants from the stains or contaminants in
the immersion oil
- Artefacts
that move only when the eyepiece is rotated are in the eyepiece
or on its lenses. Artefacts may also be caused by material
on the condenser lenses, mirror, or light source
- Keep
all the slides for external quality control, according to
established procedures (see I. Organisation and management).
- Analyse
results on a weekly and monthly basis for percentage of positive
results. Investigate any sharp differences from the
norm
Morphological
characteristics of acid-fast bacilli
- Acid-fast
bacilli are approximately 1-10Fm long and typically
appear as slender, rod-shaped bacilli, but they also may appear
curved or bent
- With
carbolfuchsin staining, tubercle bacilli look like fine red
rods, slightly curved, more or less granular, isolated, in
pairs or in groups, standing out clearly against the
blue background
- Individual
bacteria may display heavily stained areas referred
to as ?beads?
and areas of alternating stain may produce a banded
appearance
- With
fluorochrome staining, tubercle bacilli are rod shaped
and emit a bright yellow fluorescence against a pale yellow
(potassium permanganate) or orange (acridine orange)
background
- Some
mycobacteria other than M. tuberculosis may appear
pleomorphic, ranging in appearance from long rods to coccoid
forms, with more uniform distribution of staining properties
- Organisms
other than mycobacteria may demonstrate various degrees
of acid-fastness. Such organisms include Rhodococcus
spp., Nocardia spp., Legionella spp., and the
cysts of Cryptosporidium and Isospora spp.
- Rapidly
growing mycobacteria may vary in their abilities to
retain acid-fast stains
Causes of error
in microscopy
Connected
with the specimen
- Inadequate
specimen quality and/or volume
- Inefficient
washing of re-usable containers may leave residual bacilli
which give rise to false positives
- Carelessness
in marking the container. Marking should be done on the body
of the container and not on its lid
Connected
with the preparation of the smear
- Insufficient
or poorly lit work surface
- Mixing-up
of slides
- Preparing
too many slides at once. A maximum of 12 is recommended
- Utilisation
of slides that have been positive. These should be discarded
- Specimen
contamination due to careless use of loops, pipettes or wooden
applicators
Connected
with the staining
- Using
scratched slides on which deposits of stain may look like
bacilli
- Using
unfiltered fuchsin which may contain crystals
- Carelessness
in heating the fuchsin, allowing it to dry and crystallise
on the smear
- Inadequate
decolourising of the smear which may leave red stain
on saprophytic bacilli which then appear to be acid-fast
Connected
with the microscopic examination
- Failure
to check the slides and renumber them should the number be
obscured during staining. This may lead to substitutions
- Failure
to clean the immersion lens with lens tissue after each examination,
especially after a smear was found to be positive
- Presence of bacilli in the
immersion oil owing to the practice of touching the smear
with the neck of the bottle. A dropper bottle should be used
and the drop allowed to fall without any contact between
bottle and slide
- Erroneous
recording of the results
Troubleshooting
guide for microscopy
| Problem |
Possible
causes |
Solution |
| Field
is dim |
Condenser
may be too low
Condenser iris may be closed |
Raise
the condenser
Open the diaphragm |
|
Dark
shadows in the field which move when eye piece is
moved |
Eye
piece may be dirty
Eye piece or objective may be contaminated with fungi
Surface of eye piece may be scratched |
Clean
the eye piece
Eye piece may need repair
A new eye piece may be needed |
| The
image is not clear |
The
smeared portion of the slide may be upside down
There may be an air bubble in the oil
The oil may be of poor quality
There may be dirt on the lens |
Turn
the slide over
Move the 100x lens from side to side
Use only good quality immersion oil
Clean the lens |
|
The
image through low power is not clear |
There
may be oil on the lens
There may be dust on the upper surface of the lens
The lens may be broken |
Clean
the lens
Clean the lens
A new lens may be needed |
Consequences
of false-positive and false-negative smears
False-positive
results
- Patients
are started on treatment unnecessarily
- Tuberculosis medications
are wasted
- In
follow-up examinations the intensive phase of treatment is
continued longer than necessary
- Patients
may loose confidence in the National Tuberculosis Programme
False-negative results
- Patients
with tuberculosis are not treated, resulting in suffering,
spread of tuberculosis and death
- Intensive
phase treatment is not extended for the required duration,
resulting in inadequate treatment
- Patients may loose
confidence in the National Tuberculosis Programme
|